Temperature Range
0 to 1000°C
Temperature Output
Voltage: 0-5 V DC or 0-10 V DC output, linear with measured temperature, rescalable
Alarm Output
Open collector alarm output with temperature threshold and hysteresis, configurable
Note: Alarm and temperature outputs may be used simultaneously
Field of View
15:1 (see Field of View Diagram)
Accuracy
± 1.5% of reading or ± 1.5°C, whichever is greater
Repeatability
± 0.5% of reading or ± 0.5°C, whichever is greater
Response Time
125 ms (90% response)
Configuration
Via built-in display and controls
Emissivity Setting
Adjustable 0.2 to 1.0
Max Temperature Span (Voltage Output)
1000°C
Min Temperature Span (Voltage Output)
100°C
Spectral Range
8-14 μm
Max. Supply Voltage
28 V DC
Min. Supply Voltage (at Sensor)
12 V DC (for 10 V output)
6 V DC (for 5 V output)
Max Current Draw:
6 mA
Configurable Parameters
Temperature range for voltage output
Output voltage range and scale
Alarm output threshold and hysteresis
Emissivity setting
Reflected temperature
Signal processing (see below)
Temperature Units
°C / °F
Signal Processing
Averaging Period (0.125 to 60 seconds)
Peak / Valley Hold Mode
Hold Period (0.125 to 120 seconds)